Firmware versions:

medilog AR: Firmware V3.1, V3.0 was released with Darwin 2.10.1

AR firmware history:

  • V3.1 fixes a bug concerning the switch-off behaviour as soon as an external battery is used. With version 3.0, switching off the recorder does not work as expected. The medilogAR keeps switching on by itself as soon as the external battery is inserted, which causes the external and internal battery of the recorder to be discharged. This bug only affects recorders with serial numbers 10000 or higher.
  • V3.0 added support for BTLE based recorders
  • V2.16 introduced OfficeU and added 48h profiles.
  • V2.14 finally resolves problem, that some recorders show interrupted recordings. Also improves signal quality. Requires Darwin2.9.2. To update the recorder,  install "Darwin_Firmware_AR_V214.exe" on top of Darwin and update the recorder with Darwin. You only need to install the patch on the PCs where recorders are configured. The next time you try to configure a recorder via USB, the configuration tool will recognize the old firmware of the recorder and update it automatically.
  • V2.1: addresses issue that leads to interruptions in ECG
  • V2.0 added integrity verification for persistent SD card memory. Requires at least Darwin 2.9.2 
  • V1.21 reduced 72 hour recordings that stopped too early.

AR4plus, AR12plus, FD5plus: HW/SW: 6.5 / 3.20

FD12plus: 1.20

BR-102plus: V2.10s1, V1.21, V1.16