Not particularly Darwin, but just a few facts on the medilog recorders:

Reason for end of recording

Start Darwin2 and open the recording. Hover the mouse over the recording length. A hint text shows the end time and the reason why the recording has ended.

Changing cable types:

* FD12plus: When changing the cable (e.g. from 4 to 10 lead), you need to "inform" the recorder with a special ini-file on the SD card (ask fot the "Change calbe type of FD12 plus.zip" file, I can send it to you via email)

* For the "smaller" recorders (e.g. AR12plus), you also need a new firmware for a new cable type.

Current firmware versions

Recorder error codes

  • AR12plus: "key" + 2 : CRC of internal configuration data sector corrupt: Please send the serial number and and the cable configuration. We send back repair data
  • MT101: "Unexpected returncode: 1500" : No data found. Please try to take the memory card out of the recorder and import the recording directly by inserting the card into a card reader.

Summer/wintertime

medilog recorders perform the change between summer and wintertime automatically. medilog AR recorders sync with the PC time whenever you configure the recorder using the USB connection. medilog ARxplus recorders that are configured using an SD card always get the info in which time relative to GMT0 the PC "lives". (e.g. in Vienna in summer you'll get the value "+1" here). When this time difference changes (e.g. in Winter in Vienna -> "+2"), the recorder knows that the time was changed, and changes the internal clock accordingly.

1000 Hz recordings fail

If you have a recorder that fails to record at 1000 Hz, make sure you use Cactus 2GB industrial grad cards (with Adapter) that is also shipped with the medilogAR recorder. 1000 Hz recordings require a very fast response from the card and the Sandisk cards fail in that case.