7 day recordings:

Using an internal and an exchangeable battery the upcoming medilog AR recorder will be capable of recording up to 14 days of ECG (Infinity mode). If you then charge the internal battery after 2 weeks on a wall charger USB for 2,5 hours, it will continue to recorder another week. Other recorders may require a battery change from time to time. This normally opposes no problem, as the patient needs to place new electrodes regularly anyway.

Battery lifetime:

  • FD12plus with a normal battery ~ 40 hours
  • FD12plus with Lithium energizer battery approx 8 days

Changing the battery:

* Battery change in infinite mode: after inserting a new battery a screen pops up indicating that there's an "old" recording. If you do nothing the old recording is continued.

* Battery change in non-infinite mode: Time for battery change is max 30 sec. After 30 seconds the recording is considered as ENDED. When the recorder is switched on after a potential battery change, a screen will indicate that there's a recording. (you'd need to press both buttons to delete it)

* In either way you MAY NOT switch off the recorder to change the battery. This would end the recording.