Network security
Connections to the medilog® DARWIN2 database and HIS connection (if used) via HL7 or similar use network communication.
Some communication standards like HL7 do not offer encryption. If a software connects to medilog® DARWIN2 using such a standard, the communication will happen in plain text over the network. Additionally, if a Firebird version older than 4.0 is used, the communication with the database remains unencrypted.
To prevent attackers from getting access to sensitive data, the network needs to be secured. Recommended measures are:
- Restrict physical access to the network.
- Use a firewall, so that only medilog® DARWIN2 workstations can connect to the firebird database.
- If you cannot restrict physical access to the network, you may want to use encryption/tunnelling for potentially communication paths.
Explanation: An attacker who manages to plug into your network can use a network sniffing software to listen to all network traffic and find unencrypted HL7 messages that contain sensitive patient data. By setting up a secure tunnel, you can establish an encrypted communication path between two fixed points in the network. Communication between these points (e.g. the HL7 caller on one side, medilog® DARWIN2 on the other side) is secure and cannot be intercepted.
See also Firebird user and password
Restricting access to the PC
Attackers may try to install malicious software on servers/workstations to gain access to passwords or other sensitive data. To prevent this, restrict normal users from installing software.
Attackers may try to copy the Firebird database file to gain access to the stored information. To prevent this, restrict access to the location where the database file is stored. (You can remove read-access-permissions for all users for the folder where the Firebird database is stored as medilog® Darwin2 accesses the database trough a system service)
Attackers may try to delete data from the shared raw data folder. Restrict access to the shared raw data folder.
Attackers may try to read or delete archive files from your archive drive. Restrict physical and digital access to the archive drive.
Audit log
From Version2.10 on Darwin2 logs various events (Failed login attempt, deleting a recording, report generation) in the Windows event service (Event source: "MedilogDarwin"). Darwin2 can be configured, so that all workstations send these messages to a central event server. Specialized IDS/IPS HIDS software systems (e.g. SIEM https://www.solarwinds.com/en/security-event-manager) that monitor these logs can be used to automatically detect potential cybersecurity attacks.
- Details
- Written by: medilog
- Category: Install Darwin2
- Hits: 760
This is a guideline how to test updating to a new version of Darwin directly in a productive environment.
ATT: If we are looking at a network installation with more than one workstation: While the test is running, the "normal" users will not be able to start the old version of Darwin, as the license server can only be bound to ONE database at a time. (see point 5 of chapter "Revert to original installation")
We will create a copy of the existing database & settings, install the update, do tests and in the end revert to the old version.
For this we need installers of the old Darwin version that currently is in use and the new installer. We only need the Darwin-installer (05_Darwin_V2.x.x.x.exe, we will not update the database driver nor the dongle driver).
Here's what you do:
Create a clone for testing:
- If Liberty is installed on that PC: Stop the liberty service.
- Close all currently running Darwin applications (Darwin, Observer). Start the license browser and check that no other users are connected. Close the license browser afterwards.
- Open the AdminTool, switch to the Backup tab, click on database/backup. Create a folder on the desktop "test", untick "reports" and create a backup into that folder.
- Open the network-tab of the AdminTool. Note down the exact name of the database and the location of the license server on a piece of paper.
- Click the pencil symbol in the edit field for the database name. Enter a name for a Darwin database, e.g. C:\ProgramData\Darwin2\Evolution2_NewVersionTest.fdb". Click OK. Darwin will create a new database.
- Switch to the Backup tab, click on database/restore and select your backup folder. After restoring the database the AdminTool will close.
- Start the AdminTool again and change the raw data path to a new local folder. Eg your old folder was "C:\ProgramData\Darwin2\Recordings\" - change it to "C:\ProgramData\Darwin2\Recordings_NewVersionTest\" - naturally you'll have to create that folder in the windows explorer. Ignore the warning shown by the AdminTool ("...already recordings in the database..") and close it.
- Try starting Darwin. If Darwin reports, that the license is bound to a different database name, you need to resolve that: Open the license browser. Hold down the CTRL+Shift on the keyboard and right-click into the license browser. Select "Unbind" from the context menu.
- To make sure everything worked so far: Open Darwin and try to open a recording. You have to get an error message, that the recording folder couldn't be found. That's fine that way.
- Import a recording (any demo recording should do). Verify that you can open that recording in Darwin.
- Install the new version of Darwin.
Perform tests
You can now do tests with the new version (importing recordings, HL7 requests, worklist handling), but bear in mind that all those actions will be discarded at the end of your test. When you are ready with your tests, you have to...
Revert to the original installation
- Uninstall Darwin.
- Install the old version again.
- At the end of the installation process you are asked for the database location. Enter the location you noted from the original installation.
- After the admin tool connected to the database, make sure the license server location is also correct.
- Unbind the license server: Repeat point 8 of the installation-checklist above.
- Open a recording from the database to see that everything works fine.
- Manually delete the test database and the test-folder for the test raw data.
That's it. :)
- Details
- Written by: medilog
- Category: Install Darwin2
- Hits: 1706
The update is first installed on the server, then on the workstations.
For a single-seat installation (No client-server setup) only perform the steps listed in “On the server”.
On the server:
1.) Make sure all instances of Darwin applications (e.g. Darwin, Observer, DConnect) – also on network connected workstations – are closed.
2.) Make a backup of the recordings database
See chapter “Backup/restore database” of the IfU
3.) Install the new version of Darwin
Insert a CD / USB stick with the new version of medilog Darwin. The installation program should automatically start. If nothing happens manually start the application "Autorun.exe" on the installation medium.
The installer automatically detects which parts of your installation can be updated to a newer version. The appropriate buttons will be highlighted green indicating that these parts will be installed. You can overrule the selection by clicking on the appropriate buttons.
To start the installation process click "install".
4.) Test Darwin
Open Darwin. Open a recording. Verify the application starts and loads as expected.
5.) (optionally) After the installation finished update Liberty
Only if Liberty is installed.
On the workstations:
On workstations you only need to install the new version of Darwin on top of the old one.
You don’t need to update/install database drivers, perform backups etc.
- Details
- Written by: medilog
- Category: Install Darwin2
- Hits: 1502
If you are upgrading an existing Darwin1 customer to Darwin2 we suggest to leave Darwin1 on the PC and install Darwin2 on the same machine.
You can run both versions of Darwin in parallel on the same machine. However Darwin1 and Darwin2 can not share the same database. So if you want to move Darwin1 recordings to a Darwin2 database you will have to do that via DAW-files (export from Darwin1 to DAW and import those files into Darwin2). However, we recommend to leave Darwin1 on the computer. If a customer then wants to look into an old recording, he can simply open Darwin1 to do so.
Old configuration: Darwin2 has undergone a lot of functional changes. Unfortunately it's not possible to re-use screen layouts or reports that have been created in Darwin1. Also the DConnect-configuration has to be re-done (see below)
ATTENTION: When you import a DAW file that was created with Darwin1 into Darwin2 the results of the beat analysis and the arrhythmia analysis are NOT imported. To get beat annotations you will need to re-analyze the recording. (Reason for that is that the changes in the beat analysis structure were so big, an importing routine would have been an enormous effort).
Using old dongles: Darwin2 will work with Darwin1 dongles, unless they are VERY old (older than 2006). Old dongles are purple or blue and have a rather round shape where newer dongles are black and have a square shape. When you installed Darwin2 on a machine you can run the tool named "DongleTest.exe" that is installed in the Darwin2 program folder to check if your dongle is new enough to support Darwin2. If the dongle is too old please contact Switzerland for a dongle exchange. You can exchange the old dongle against a SoftLicense for a small handling fee.

Known issues:
64bit Firebird driver: If you install Darwin2 on a 64bit Windows, then also the 64bit driver version of the Firebird database driver will be installed. Unfortunately Darwin1 can't work with that version. To fix that please uninstall all installed firebird drivers and install the 32bit firebird driver version (on the Darwin2 installation CD, folder "Supplements", file "02_Firebird-2.5.???_Win32.exe"). From Darwin2.1 on the installation routine automatically forces installation of the 32bit driver if Darwin1 is detected on the system.
Darwin1 license server: To use Darwin1 and Darwin2 in parallel, Darwin1 will need to be at least in version 1.12 so it uses the license server.
Upgrading DConnect (HL7-connectivity): Switching to the new Darwin2 DConnect service is unfortunately not that straightforward. From the old folder where Darwin1 is installed collect the following files and mail them to us: DConnect_custom.ini, Tpl_*.txt, HL7*.txt
We will tell you how to set up the new system to behave like the old one.
- Details
- Written by: medilog
- Category: Install Darwin2
- Hits: 3142
- multiple clients can connect to a common database
- multiple clients share licenses, provided by a license server