Darwin calculates various values, eg:
HRV Tabsum:
- SDNN [ms]: Standard deviation of RR-intervals (according to HRV data prep, normally only N-beats)
- SDNN-i [ms]: average of all standard deviations of the time windows (the recording is split into defined time windows and then the standard deviation of the RR intervals for each time window is calculated. SDNN-i then shows the mean value of all the calculated standard deviations).
- SDANN-i [ms]: standard deviation of all mean values of the time windows (the recording again is split into time windows but then the mean value of each time window is calculated. SDANN-i shows the standard deviation of these mean values).
- Circadian index: ratio between the average HR during the night and during the day
- HRV index: heart rate variability triangular index. This is a geometric measurement giving the total number of intervals divided by the height of the histogram of all NN intervals
- TINN: A geometric measurement of the Triangular index of the NN time distribution (baseline width). That is baseline width of triangular interpolation of the highest peak of the histogram of all NN intervals
- r-MSSD [ms]: square root of the average of the squared differences between consecutive RR intervals
- pNN50 [%]: percentage of intervals that differ more than 50 ms from the preceding interval
A (german) web page explaining some of these values can be found here.