medilog® DARWIN2 offers a method to automatically archive recordings that have been released and not been accessed for a specified period of time.
In the Admin Tool, in the extended settings, section "Recording database", you will find the following settings:
- Auto-archive age threshold
- Archive path
- Discard when archiving (optional)
- Delete DAW files after x days
- Permanently delete recordings after x days
Process:
- Auto-archive age threshold: All recordings that are marked as released and have not been opened for more than the specified number of days will be archived to the path given in “Archive path”. If “Discard when archiving” is set, those recordings are not archived, but only the raw data is deleted and the PDF remains.
- Delete DAW files after x days: archived recordings are deleted after the specified number of days. The database entries and the associated PDFs will remain available.
- Permanently delete recordings after x days: also the database entries and the PDFs are deleted after the specified threshold.
To make sure that auto archiving is possible, the application "ArchiveDaemon.exe" needs to be run on one of the workstations from time to time. This application can be put it into the Start menu. It is recommended to launch it from a Windows scheduler, ideally directly on the server and during night time when network usage is not an issue. To run the application manually, type “ArchiveDaemon” into the Search field of the Windows Start menu.
Note that the current Administrator password is backed up together with the database. Hence if you restore the database at a later time, the password which was valid during the backup will be restored automatically. If you are going to restore the ECG raw data as well as the database, make sure that the database and the raw data are from an identical backup.
Restoring the database must only be completed if no user is logged on. If need be, disconnect the network to avoid any unwanted login by other users.
During the database restore process, ensure that the Observer application is not running as background process. The Observer icon is shown in the Windows task bar if the application is running. If necessary, close the application with a right-click on the icon. Then select the option Close from the context menu.
Attention new in V2.11.0:
When started without any parameter, ArchiveDaemon will only list all actions it would have performed using the current configuration. It will not undertake any actual actions. This way you can test your configuration. To tell ArchiveDaemon to really perform the actions, call it with the command line paramter "-Execute"