Darwin scans the beats for a number of arrhythmias, starting with the most severe arrhythmia, ending with the least dangerous one. Generally arrhythmias can't overlap, with only a few exceptions, that are explained below. That means that if a "dangerous" arrhythmia is found, no other arrhythmia is found at the same place.
The "priority list" of arrhythmias is:
- Pause
- Bradycardia
- VTachy & NSVT
- IVR
- VTriplet
- VCouplet
- Bigeminus
- Trigeminus
- VE
- Unused V-beats
- BBB
- PSVT
- Tachycardia
- Irregular rhythm (May overlap with all atrial arrhyhtmias, except AFib and SVEs)
- AFib (May overlap with all atrial arrhyhtmias, except IrregRh and SVEs)
- SVE's (may overlap with Tachy/Brady/MinHR/MaxHR)
- Min/Max HR
A decision tree how arrhythmias are found can be found in the Darwin manual in chapter 4.6.1 "Information on arrhythmias definitions".
To answer a frequent question: Just from the number of arrhythmias you cannot calculate the number of ventricular beats in the recording, as several ventricular-related arrhythmias can have any number of V-beats. Eg an NSVT can have only 4 but also 64 ventricular beats.
Bradycardia
Bradycardia is defined as a heart rate below the normal sinus rate. Only normal beats allowed.
Conditions for onset:
- Heart rate measured over 3 R-R intervals has to be less than BradyStartBPM (preset: 45) bpm, and each of the 3 intervals must meet this.
- Over 3 intervals (3-RR moving average) the heart rate has to drop by at least MinChange (preset: 12%).
Tachycardia/PSVT
Tachycardia is defined as a heart rate above the normal sinus rate - only normal beats are allowed. The difference to a paroxysmal supraventricular tachycardia is a sudden increase in the heart rate.
Conditions:
- Interval must be lower than the interval calculated from the start threshold (preset 130bpm = 462ms)
- Only normal beats are allowed
- The arrhythmia is terminated if the calculated average heart rate falls below the stop threshold
- The arrhythmia is recognized only if a minimum duration of at least a few intervals is met. This is configurable - the preset is 5 RR intervals.
For the PSVT there are similar conditions but there must be a certain (preset 45%) sudden increas in the heart rate (preset 110bpm) that needs to last at least for 3 RR intervals. The arrhythmia is terminated after the hr falls again below a certain threshold (preset 100bpm)
VTachy / NSVT
Ventricular Tachycardia, IVR and Salvo all describe a run of consecutive ventricular beats, the distinctions being in the number of beats and maximum rate during the run.
4 or more successive ventricular beats sandwiched between non-ventricular beats are called VT (or, if the event is shorter than 30 seconds, a NSVT). Some users define a small run of consecutive ventricular beats as a Salvo. In this case a higher number than the salvo limit would be defined as a ventricular tachycardia.
If the RR intervals between the qualifying V beats are below a threshold then this event is called an ideoventricular rhythm. (handled separately)
AFib (only works if P-wave is available)
The AFib detection uses a sliding-window calculation of 40 beats (adjustable) to apply a robust (with outlier detection) calculation of the PR-interval variance. When the PR-variance exceeds a certain value AND the RR-variance exceeds also a threshold (depending on the heart rate) then the center of the sliding window is considered to be part of an AFib. Then the sliding window is moved further to find the end of the AFib. When this is finished, a second algorithm refines the exact start/end of the AFib-section.
Configuration: The sensitivity adjusts some parameters of the outlier detection and the thresholds of the PR- and RR-variance. The window length does NOT define the minimum length of detected algorithms. It defines the number of intervals used to do statistical analysises on to detect AFib On- and Offsets. If you set lower values for the window width, you will get more and shorter periods of AFib, but more False-positive sections, if you set higher values, then the predictivity will increase but you will get more false-negative sections. An interval length of 40 seems to be a good compromise.
The algorithm was verified and tested against the MIT- and a internal test database.
Min/Max HR
The heartrate trend displayed in Darwin (module "Range viewer", signal "HR") is calculated beat-to-beat. In the module you can select for a smoothened display, but this setting doesn't have any effect on any other value in the software.
The HR-values shown in the strip directory or the tabular summary are using a moving average of the RR-intervals over typ. 3 intervals (can be configured in arr settings), beat types are completely ignored here (artifact beats are naturally ignored).
ATT: The shown HR is calculated from the average interval length and is NOT the average of "single beat heart rates". In a formula it's HR = 60 / Mean-RR-Interval-over-x-beats. Or if you know, that in i seconds there were x beats, then the HR = 60*x/i [bpm]. ATT: The shown HR is NOT the average of the the contained HR-values!! So the assumption HR = 60/( (Int1+Int2+Int3)/3 ) = (HR1+HR2+HR3) / 3 is WRONG!! That's mathematically not he same.
The minimum and maximum HR events shown as strips in the strip directory use that calculation. Typically sections where arrhythmias (except AFib, AFL, AV1, AV2.1, AV2.2, AV3, Irreg. Rh, Tachy, Brady, User and BBB) were found, are excluded. If you select "allow arrhythmias", Min/Max HR strips can occur everywhere, so e.g. also during a VT or a Salvo.
SVES
Prem = 1-Prematurity/100. A prematurity value of 30% gives a value of 0.7 here. A beat is marked as SVES if it's beat interval qualifies:
AktRR <= AvgRR*Prematurity
Changing arrhythmia settings
When you change the arrhythmia analysis settings in Darwin, you normally just change that for the single recording you just opened. If you want to make the settings the new default settings for all recordings, click on the "Defaults" button and select "Save as system default".
ATT: When you do that, all future recordings that have not yet been imported will use the new setting. If you open a recording that was analyzed before you changed the default settings, it will continue to use its old settings. If you want to use the new settings on the recording, open the arrhythmia configuration dialog, click "Defaults" and select "Reset to system defaults".
When you select "Reset to factory defaults", your current and all future recordings will use the factory default arrhythmia analysis settings.
ATT: If the menu entry "Reset to factory defaults" is disabled, your current user misses the permission "Change system default". You can find this setting in the Admin tool under "Users" -> "Customize" (of the relevant group) -> "Change system default".