When Darwin finds pauses but you can clearly see QRS complexes, there's a common reason for that:

You might have deleted beats causing "gaps" in the ECG. These gaps are interpreted as pauses.

To verify open the template editor. Fold out the menu on the left and select "Show artefacts".Go back to the location where the wrong pause was found. Do you see "X" marks in the beat display? These "X" signs denote deleted beats.

Read the article Deleting beats lead to artefacts to find out how to solve this problem.