Symptom: You have a recording that shows beautiful, clear ECG complexes, but Darwin doesn't find any beats.
Looking more detailed into the recording, you see, that long portions of the recording are inhibited (they show a red background color).
Background: The medilog recorders perform some signal processing and store that information to make the analysis by Darwin more precise. Under certain circumstances it may happen, that the recorder excludes long parts of the recording (or even the entire recording), even though there's clearly an ECG signal.
This may come from the ECG having a very low amplitude, so that the recorder thinks there is nothing to look at, or the amplitude is extremely high, so the recorder thinks it's recording noise.
Solution: You just need to tell Darwin to delete the inhibitions made by the recorder. Open the recording, switch to a screen with the ECG Detail viewer. Fold out the left side menu. Right click on an ECG channel you want to "repair". From the context menu select "Remove all inhibitions". Do that for all ECG channels. Then re-analyze the recording.
Voila!
Have fun :)