The Template Editor is one of the most important parts of the software as a typical user spends most time in this module to review and correct the results of the beat analysis.

  • Open a recording, switch to the “Templates” layout
  • If the recording has not yet been analyzed - >fold out the module menu (by clicking on the vertical module caption on the left border of the module) and press the “Analyze” button.

Settings

When you change any of the settings below you need to click the "Reanalyze" button afterwards for the change to take effect.

  • Low amplitude ECG: this setting is only recommended for patients in which ECG signal amplitude is low. If this option is set and the signals are reanalysed, a more sensitive analysis algorithm is used, allowing to detect even low-amplitude QRS complexes, but also increasing the risk of muscle artefacts and noise being detected as R peaks.
  • Analyze channels: Unselect channels with a very bad signal quality to ignore these channels during the analysis.
  • Narrow PM spikes: When analyzing a pacemaker measurement, the area around pacemaker pulses are inhibited for beat detection, so in a certain period of time around the pacemaker spikes certainly no beats are found. With this setting you can tell the beat analysis that the PM spikes are "negligible" small and the whole ECG should be examined.

Display settings

  • Show artefacts: templates classified as artefacts are displayed. This setting also affects the ECG Detail Viewer Module.

Reclassifying templates

To reclassify a template, you can highlight it in the main screen and use the icons at the bottom to assign another classification:

  • N: normal
  • V: ventricular
  • A: Abarrant
  • BBB: bundle-branch block
  • J: junctional
  • X: artefact (artefacts are excluded from the analysis)
  • Z: breaking artefact (act like short exclusions, so on the same position no arrhythmias like Pauses or bradycardias will be detected.
  • U1 - U4: defined by the user: to define this classification, right-click the icon and select “Edit user-defined templates”. Select the template, enter/change the name and define the classification.

For pacemaker recordings additionally the following beat types are possible:

  • P: paced (only displayed for pacemaker recordings)
  • Pa: atrial paced
  • Pv: ventricular paced
  • Pdc: dual-chamber paced
  • Pf: fusion paced

 

Working in the waterfall view

  • Templates are sorted by the number of contained beats. Normally the 1st template is perfect. Quickly scroll through the beats in the waterfall view. Mark the template so you see that you have worked through it.
  • Jump through the templates and mark those that are perfectly clean as edited.
  • If the displayed ECG signal is noisy (e.g. an electrode fell off) you should check a different ECG-channel. To switch between channels use the context menu of the waterfall view of hold down the ALT-key and roll the mouse wheel.
  • If the signal has a vewy high or a very low amplitude you can zoom in/out by holding the SHIFT key and roll the mouse wheel.
  • If you zoom in on the ECG if might occur that the ECG-lines overlap or come too close to each other. Simply grab the blue zoom slider and adjust the waterfall line distance by moving the slider up and down.
  • Jump from beat to beat by holding down the CTRL key and roll the mouse wheel.