When configuring recorders, there are a few things to know:
Query patient demographics: You can fill in the field "Patient ID" and click the Patient-icon. From a dropdown menu you can then select to get patient demographics for that patient ID from the database, via HL7 (PDQ patient query) or from Sema3. To speed up that process, open the AdminTool. In the extended settings under "Recorder Configurator"/"Auto-query patient data" select e.g. "via Sema3". If you now enter a patient ID and then press CTRL+Enter, then RecConfig will immediately try to get the demographics from Sema3.
Clean up Ref.Doctor names list: The field "Referring doctor" uses an intelligent list in the dropdown menu. The last 20 entries are stored in the registry. In Darwin versions before V2.9, you can only modify the list in the main application (see Recording Information Module). From V2.9 on you can delete an entry from the list by selecting the entry and then pressing CTRL+delete.
Setting up the medilogAR
Click the icon Recorder setup (see Database screen in IfU) to configure a recorder. Enter the patient details or click the picture to select a patient from the database. Then select the recorder at the bottom. If your recorder is not shown in this list, it needs to be activated in the Admin tool (see Setting local components in IfU). The dialogue shown on the left is displayed. Select your preferred setup option: memory card, USB or Bluetooth - see below for more information on these options.
Performing settings (common to all setup methods)
- Recording profile: select recorder profiles optimized for recordings with a duration of 24, 48, 72 hours or open end recording (=as long as possible), or Scientific (=as detailed as possible). For more information on the profiles, see the medilog®AR user guide. ATT: The profile does NOT force the recorder to power down after a specific time.
- Activate or deactivate pacemaker detection: when this option is activated, the type of pacemaker needs to be specified.
- Extended settings:
- LED flashing: a blinking LED indicates an ongoing recording
- Auto power-off: Recorder powers off after the time indicated by the profile name
- Tap-detection: Users can set markers not only by pressing the buttons on the recorder, but also by tapping the recorder firmly three times. DARWIN2 detects these events when uploading and sets the corresponding marks in the patient diary.
- When clicking the green arrow icon, the patient data is written to the memory card. Instructions are given on how to start the recording and where to place the electrodes. Moreover, you can print a patient diary template. See recorder user guide for more information.
Configuration with memory card
Click the Star icon to select the recorder version (Office, Professional or Enterprise
Configuration via USB
The recorder edition, (Office, Professional, Enterprise), the serial number and battery state is displayed. Click the drop-down list to select from all available recorders. Click the star icon to start configuration of the selected recorder
Setting up the BR102plus
- The recording profile can only be set, when the firmware of the attached recorder allows that.
- Baulmann profile: PWA is measured only at the beginning of the recording four
times in 5-minute intervals