HRV data preparation
This module is used to configure which data is used for the HRV data analysis.
By default only N-beat are considered for the HRV analysis. Recordings with a large portion of ventricular beats will show gaps in HRV trends and -tables. By unticking "Only normal beats", also ventricular beats will be used for the calculation.
When you change values, the preview gives an impression of the amount of data that an be used for the calculation. As a final step click "Apply" to apply your settings.
HRV Tabular summary
Gives a tabular overview over statistical values, such as
- SDNN (ms) – Standard deviation of all NN intervals usually over 24 hours.
- SDANN (ms) – Standard deviation of the averages of NN intervals in all 5-minute segments of the
entire (24-hour) recording. - RMSSD (ms) – The square root of the mean of the sum of the squares of differences between
adjacent NN intervals. - SDNN index (ms) – Mean of the standard deviations of all NN intervals for all 5-minute segments
of the entire (24-hour) recording - pNN50: Percentage of adjacent NN differing by more than 50ms